TSI AeroTrak 9306-V2 Particle Counter Rental
- Particle counter with 6 programmable channel sizes between 0.3 μm to 10.0 μm (max measurable particle size is 25μm).
- Modes: differential (counts particles between channel sizes), cumulative (counts particles greater than each channel size), and concentration (ft3 or m3 concentration).
- Rechargeable Li-ion battery provides approx 7 hours of continuous use.
- 0.1 CFM (2.83 L/min) sampling rate.
- Datalog 10,000 sample records: includes date, time, six particle channels, flow, ID, and sample volume.
Disclaimer: The TSI AeroTrak is NOT intended for use when conducting total, respirable, PM2.5, or PM10 dust monitoring. Such use can damage the instrument. Please see the Casella CEL-712 Dust Monitor Rental for this application.
The TSI AeroTrak 9306-V2 is a handheld particle counter that uses a Class I laser. The instrument features 6 channels that can be programmed to sample air particle ranges from 0.3 μm to 10.0 μm. The built in pump on the has a flow rate of 0.1 CFM (2.83 LPM). The particle counter is ISO 21501-4, 21 CFR 1040.10 and 21 CFR 1040.11 complaint for light-emitting products. The AeroTrak uses a rechargeable lithium-ion battery with a battery life of up to 7 hours of continuous use. The rental instrument recieves an annual factory service and comes with certificates of calibration.
Typical applications for this particle counter include clean room monitoring, research, exposure assessment, indoor air quality, filter testing, clearance testing, quality assurance, and contaminant migration studies. All AEROTRAK™ particle counters meet JIS standards.
Size Range | 0.3 to 25 µm |
---|---|
Channel Sizes | Standard: 0.3 to 10 μm, user-selectable; factory calibrated at 0.3, 0.5, 1.0, 3.0, 5.0, 10.0 μm. Additional channel sizes available |
Size Resolution | <15% at 0.5 μm (per ISO 21501-4 requirements) |
Counting Efficiency | 50% at 0.3 μm; 100% for particles > 0.45 μm (per JIS and ISO 21501-4) |
Concentration Limits | 3,000,000 particles/ft3 at 5% coincidence loss |
Light Source | Long life laser diode |
Zero Count Level | <1 count/5 minutes (per JIS B9921 and ISO 21501-4) |
Flow Rate | 0.1 CFM (2.83 L/min) with ±5% accuracy (meets JIS and ISO 21501-4 requirements) |
Calibration | NIST traceable with TSI calibration system |
Calibration Frequency | Recommended minimum once per year |
Sample Probe/Tubing | Isokinetic sampling probe |
Sampling Modes | Manual, automatic, beep, cumulative/differential count or concentration |
Sampling Time | 1 second to 99 hours |
Sampling Frequency | 1 to 9999 cycles or continuous |
Exhaust | Internally filtered |
Vacuum Source | Internal pump |
Communication Mode | Modbus® TCP over Ethernet or USB |
Data Storage | 10,000 sample records: includes date, time, six particle channels, flow, ID, and sample volume; transferable via USB data download or TRAKPRO™ Lite software |
Data Security | Password protected |
Alarm/Status | Audible alarm on counts, low battery, and sensor status indicators |
Display | QVGA 3.5-inch (8.9-cm) touch screen with Windows® CE operating system |
Languages | English, Spanish, German, French, Italian, Japanese, and Chinese (simplified) |
Reports | Provides Pass/Fail on FS-209E, ISO 14644-1 and EU GMP |
External Surfaces | High impact injection molded plastic |
AC Power (power to AC adapter) | 110 to 240 VAC 50 to 60 Hz Universal in-line power supply |
DC Power (power to DC instrument) | 12 VDC @ 2.5 A |
Battery | Removable/rechargeable Li-Ion |
Battery Life | Up to 7 hours of continuous use |
Recharge Time | 4 hours |
Dimensions (L x W x H) | 9.4 x 4.9 x 3.2 in.(23.9 x 12.4 x 8.1 cm) |
Weight | 1.0 kg (2.2 lbs) with battery |
Standards | CE, JIS B9921, ISO 21501-4 as listed above |
Operating Conditions | 41 to 95°F (5°C to 35°C); 20% to 95% noncondensing relative humidity |
Storage Conditions | 32 to 122°F (0°C to 50°C); Up to 98% noncondensing relative humidity |
Included Accessories | Power supply, power cord, battery, isokinetic inlet, stylus, purge filter, TRAKPRO™ Lite data download software, operational manual on CD, computer cable, calibration certificate, and Quick Start Guide |
CE Marking |
|
RoHS Marking | Yes |
- TSI AeroTrak 9306-V2 Particle Counter
- Zero Adapter
- TrakPro Lite Software
- Owners Manual
- Battery Charger
- Factory Calibration Certificate
- USB Data Cable