  • Fully AARST-NRPP and NRSB certified radon detector for the US market.
  • Radon measurement: 0 - 2700 pCi/L and 0 - 100 000 Bq/m3.
  • Also measures temperature, relative humidity, and barometric pressure.
  • Control the meter and view the data via mobile phone application.
  • Stores up to 5 years of hourly data points.
SKU
RN-CRM-RADON
Rental Rates
Daily Weekly Monthly
$72.00 $168.00 $435.00
A fully AARST-NRPP and NRSB certified product for radon professionals. Corentium Pro features four highly accurate radon chambers that work in parallel for maximum performance, reliability and speed.

In addition to having full app support, it also can be read out using the CRA Software page. In CRA you can analyze the data in great detail and create reports suited for your use.
Power Three LR06 (AA alkaline) batteries
Battery Life Approximately 2 years
Color Black plastic with yellow front plate
Dimensions Height: 140 mm (5.5 inches)
Width: 140 mm (5.5 inches)
Thickness: 25 mm (1 inch)
Weight 10.6 ounces (including batteries) 300 grams (including batteries)
Operating Conditions Temperature: 39 to 104 °F (4 to 40 °C)
Relative humidity < 85%
Sampling Method Passive radon diffusion chamber
Alpha spectrometry
Detection Method 4 silicon photodiodes in 4 distinct radon chambers
Measurement Range 0 - 2700 pCi/L
0 - 100 000 Bq/m3
Measurement Uncertainty After 24 Hours
σ ~ 7 % ± 0.15 pCi/L
σ ~ 7 % ± 5 Bq/m3

After 7 Days
σ~ 5 % ± 0.06 pCi/L
σ ~ 5 % ± 2 Bq/m3
Diffusion Time Constant 25 minutes
Temperature Sensor Range: 39 to 104 °F (4 to 40 °C)
Resolution: 0.36 °F (0.2 °C)
Accuracy:
± 1 °F (typical), ± 2 °F (max)
± 0.5 °C (typical), ± 1 °C (max)
Humidity Sensor Range: 5% RH to 85% RH (noncondensing)
Resolution: 0.5% RH
Accuracy: ± 4.5% (in range 20 - 85% RH)
Barometric Pressure Range: 50.0 kPa to 115.0 kPa
Resolution: 0.06 kPa
Accuracy: ± 1 kPa
Memory The internal memory stores 5 years of hourly datapoints
Memory Capacity ~ 1 900 days of measurements
~ 5 datasets of one-year length
~ 177 datasets of 1-week length
~ 325 datasets of 2-day length
Memory Type Non-volatile flash memory
  • 1 Corentium Pro Digital Radon Monitor
  • 3 x AA batteries
  • User manual
  • Software