Simple operation, no special training required.
On-the-spot results
Portable, completely self-contained, weighs only 5 lb..
Accurate
Readings automatically compensated for battery voltage changes, temperature variations, and differences in altitude
Audible alarm
Activates when gas concentrations exceed user
adjustable set point or meter needle falls below zero. Also alerts
user of low battery condition
Precision meter
Easy to read parallax mirror scale with clear graduations
Long life sensor
Non reductive and resists poisoning by lead or silicone
Recorder output
Time weighted evaluation or permanent record
Rechargeable battery pack
Over 8 hours of continuous operation
Earphone operation
For areas of high ambient noise
Aggressive Fluid Vapor Recovery System Monitoring (AFVR)
VOC Detection
Gas Leak Detection
Combustible Gas Detection
| Function |
Measurement of low concentrations of flammable vapors, relative to health, safety, and pollution control standards |
| Standard Measuring Ranges |
Calibrated on Hexane: 0-100, 0-1000, and 0-10,000 ppm
With Optional Dilution probe: 0-100, 0-1000, 0-10,000, 0-100,000 ppm (0-10% by volume) |
| Accuracy & Reliability |
The extent to which meter readings correspond to actual parts per million of
combustible gases in sampled air depend upon (1) the internal electrical
stability of the instrument, (2) proper calibration of the instrument on gas
mixtures of known concentration, and (3) purity of the air sample used for
meter pointer zero setting. |
| Sensor Type |
Catalytic combustion - long-life, non reductive and resistant to poisoning by lead and silicones |
| Minimum Detectable Concentrations |
2 ppm |
| Sampling Method |
Integral pump |
| Sample Flow |
1,750 cc/minute nominal (with inline filter and hose probe assembly) |
| Recorder Output |
Range: 0-100 MV; Impedance: 10,000 ohms or greater |
| Audible Alarms |
Pulsating tone activated when user adjusted set point is exceeded |
| Battery Capacity |
Better than 8 Hours continuous operation with nickel cadmium rechargeable batteries |
| Display |
Precision analog meter with Parallax Mirror scale and clear graduations |
| Operating Temperature |
14o to 120o F (-10o to 49o C) |
| Response Time |
90% of final reading in less than 30 seconds |
| Warm-up Time |
One minute |
| Housing Construction |
Reinforced metal top, end panels and carrying handle with tough ABS Plastic wrap around side cover |
Dimensions |
9" L x 3.75" W x 6.625" H |
| Weight |
5 lb.. |
| Hazardous Area Use |
FM approved Intrinsically Safe for use in Class I, Division I, Groups C and D; and Non-Incendive, for Class I, Division 2, Groups A, B, C, and D |
|
|
Daily |
Weekly |
Monthly |
$15.00 |
$45.00 |
$135.00 |
|
Dilution Probe
|
|
Daily |
Weekly |
Monthly |
$10.00 |
$30.00 |
$70.00 |
|
|
*NOTE: All Organic Vapor Analyzers (PID, FID, TLV Sniffer) and Gas Meters are calibrated prior to each rental. If you wish to purchase calibration gas, please call for pricing.
Also, we no longer ship calibration kits for rental since the return shipper must be a certified hazardous materials shipper. Rental rates are based on 25% calibration gas usage per week (based on tank pressure).
Overages will be billed in 25% increments in addition to the rental rate.
|