Environmental Monitoring Equipment & Supplies • Sales, Rental & Service by Enviro-Equipment — Charlotte, North Carolina USA

  • Measures turbidity with a range of 0 to 1100 NTU.
  • White Light Lamp conforms with USEPA Method 180.1
  • Simple measurement and calibration
Title Daily Rate Weekly Rate Monthly Rate
HF Scientific MicroTPW Turbidimeter Rental $30.00 $90.00 $270.00

Designed to provide the ease of portability needed in the field with rugged durability, the MicroTPW is a necessity for anyone monitoring turbidity on the go.  The shock-resistant carrying case holds everything necessary for field operation while the instrument itself removes easily to go wherever it is needed.  With resolution of 0.01 NTU and an extended range to 1,100 NTU, the EPA Accepted White Light MicroTPW is a perfect partner for field use.

White Light EPA Accepted Method

conforms to US-EPA Method 180.1

Rugged Self Contained Carrying Case

Contains everything needed, including battery pack, manual and calibration standards.  Ensures durability and convenience regardless of where the sample needs to be taken.

Waterproof

Waterproof housing allows sample measuring and cleaning in any wet environment.

Auto Ranging 0 - 1,100 NTU

Instrument senses turbidity level of sample and automatically adjusts to the appropriate measurement.

Simple Calibration Procedures

Calibration initiated with the push of a button ensures accurate readings.

PrimeTime Standards Kit

Each MicroTPW Turbidimeter is shipped with an HF PrimeTime Calibration kit manufactured using USEPA Accepted Primary Standard materials.

Economical Cost

Over 5000 Tests on a single set of 4 - AAA alkaline batteries Tungsten Lamp Life 7 - 10 years Built in Diagnostics

Items Included:
  • Items Included:
  • HF Scientific MicroTPW
  • Calibration Standards
  • Calibration Record
  • 2 Measuring Cuvettes
  • Instruction Manual
  • Spare Batteries

Conformance

USEPA Method 180.1

Range

Auto ranging from 0 - 1100 NTU

Resolution

  • 0.01 NTU < 100 NTU
  • 0.1 NTU < 100 - 999.9 NTU
  • 1 NTU < 1000 - 1100 NTU

Accuracy

  • ± 2% of reading or ± 0.01 NTU (0-500 NTU)
  • ± 3% of reading (500-1100 NTU)

Response time

14 seconds

Power supply

4 - AAA alkaline batteries (over 5000 tests)

Display

4 digit (7 segment) LCD

Principle of operation

Nephelometric

Certification

CE, NEMA 4X. Designed to meet IP67

Light source

White light - (tungsten)

Sample Cells

10 ml