Environmental Monitoring Equipment & Supplies • Sales, Rental & Service by Enviro-Equipment — Charlotte, North Carolina   USA

 

Horiba W-22 XD Water Quality Monitoring System


Home » Equipment Sales » Water Quality Testing Instruments » Horiba W-22 XD Water Quality Monitoring System

Years of Horiba sensor technology development have reached their culmination in the form of a 47-mm diameter sensor probe. These units offer a compact, highly accurate monitoring solution with high pressure tolerance, long term continuous measurement capabilities, waterproof construction, and monitoring of up to 10 parameters. In addition to the Water Quality Monitoring System's own capabilities, it was designed by Horiba to be compatible with GPS and other data processing techniques.  And now, the W-22 is available with a flow through cell designed by Horiba. 

Horiba W-22 XD Water Quality Instrument


Simultaneous Measurement of up to 10 Parameters

pH, Dissolved Oxygen, Conductivity, Salinity, Total Dissolved Solids, Sea water Specific Gravity, Temperature, Turbidity, Depth, Oxygen Reduction Potential

The NEW W-22XD (Extra Durable) design

Includes a new 316 Stainless Steel Probe Cover, Stronger CT cell, and Pelican carrying case.

One Touch Calibration

Just submerge the probe into the calibration beaker filled with standard solution and press the button for simultaneous one-point calibration.  The system also enables two-point calibration where high precision measurement data is required.

Probe assembly fits inside 2-inch well

Data Logging Capability

The sensor's built in memory function enables continuous measurement for as long as one month (with measurements every 15 minutes).

Measurement at Depths up to 100 Meters

High-Precision Dissolved Oxygen Sensor

This quick response sensor requires no stirring.

System Enables GPS, PC, and Printer Capabilities

Dry Cell Battery Operated

Runs on standard alkaline batteries.

Oceans and Seas

City Sewage Water

Lakes and Marshes

Factory Drainage

Dams

Farm Water

Wells and Ground Water

Nurseries

 

Main unit
External dimension: 169(H) x 98(W) x 56(D)mm
Weight: approx. 500g
Water proof construction: IP- 67
Sensor probe
Measurement temperature:  0-55°c 
Storage temperature: - 5-60°c
Measurement depth: to 100m
Maximum probe size: 95mm 
Probe length: 430mm
Continuous use available: 30days Manual data memory (2880 items) Automatic calibration
Weight: approx. 1500 g(with 10 m cable)
pH
Two - point calibration Automatic temperature compensation 
Measuring principle: Glass electrode method
Range: pH0~14 
Resolution: 0.01pH
Repeatability: ±0.05pH
Accuracy: ±0.1pH
Dissolved oxygen
Salt correction (0 to 40 ppt / automatic) Automatic temperature compensation 
Measuring principle: Diaphragm gavanic 
Method Range: 0-19.99mg /L
Resolution: 0.01mg/L
Repeatability: ±0.1mg/L 
Accuracy: ±0.2mg/L
Conductivity
Auto range, Automatic temperature conversion)(25°C) SI units
Measuring principle: 4 AC electrode 
Method range: 0~4%
Resolution: 0.01% 
Repeatability: ±0.1% 
Accuracy: ±0.3%
Salinity
Measuring principle: Conductivity
Conversion range: 0-4%
Resolution: 0.01%
Repeatability: ±0.1%
Accuracy: ±0.3%
Total Dissolved Solids (TDS)
Conversion factor setting
Measuring principle: Conductivity
Conversion range: 0-100mg/L 
Resolution: 0.1%F.S 
Repeatability: ±2mg/L 
Accuracy: ±5mg/L
Sea water specific gravity
Display Sea water specific gravity Display °t, °0, °15
Measuring principle: Conductivity
Conversion range: 0-50°t
Resolution: 0.1°t 
Repeatability: ±2°t 
Accuracy:  ±5°t
Temperature
Measuring principle: Thermistor 
Method range: 0-55°c
Resolution: 0.01°c 
Repeatability: ±0.3°c
Accuracy: ±1.0°c
Turbidity
Unit selection Measuring principle: Penetration and scattering
Method range: 0-800 NTU
Resolution: 0.1NTU 
Repeatability: ±3%
Accuracy: ±5%
Water depth

Measuring principle:
Pressure method range 
(NTU or mg / L):0-100m
Resolution: 1m Repeatability: ±3% 
Accuracy: ±5%
Oxidization reduction potential(ORP)
Measuring principle: Platinum electrode
Method range: ±1999mV
Resolution: 1mV Repeatability:±5mV
Accuracy:  ±15mV
Notes
The accuracy rating value is obtained from measurements at an intermediate point of the standard solution after two-point calibration (at room temperature andpressure). The repeatability and accuracy rating percentages are based on the full scale (except for salinity).
  • 1: Organic solvents, strong acids, and strong alkaline solvents cannot be measured.

  • 2: The maximum depth for ion measurements are 100 m for nitric acid ion, chloride ion, fluoride ion, 15 m for calcium ion, ammonia, and 3 m for potassium ion.
Title
Description
Download
Horiba W-20 Series Manual
Operator's Manual
PDF Icon
Horiba W-20 Series Software Manual
Software Manual
PDF Icon
Horiba W-20 Series Datasheet
Datasheet
PDF Icon



New Account Setup

Locations & Contact Information
Specials || Used Equipment
Rental Options &
Delivery Details
Service/Repair & Maint.
Ask The Expert! || Blog
Sign Up For Our Newsletter!
Why Buy From Us?
Frequently Asked Questions

check out our Facebook
follow us on Twitter
see our videos on YouTube

Share This Website!

Click to Return to MAIN Home Page
Click to Return to MAIN Home Page
Click to Return to MAIN Home Page
Click to Return to MAIN Home Page
home || about us || products - general: rentalsales || products - remediation: rentalsalesused || services || order form || contact information || site map
© 2002-2012 — All Rights Reserved — Enviro-Equipment, Inc — Environmental Equipment & Supplies: Rental, Sales, Service & Repair