Environmental Monitoring Equipment Environmental Equipment & Supplies
Orders Specials Ask The Expert Blog Testimonials
header right
 


Cirrus Research doseBadge Personal Noise Dosimeter

Features
Applications
Specifications

  • Standards
  • Measurement Range
  • Measurement Functions
  • Memory
  • Weightings
  • Exchange Rate
  • doseBadge Configuration
  •  

    The doseBadge is a unique solution for Personal Noise measurements. The new CR:100B version of the doseBadge provides not only the measurements essential for compliance with Occupational Noise regulations, but with the new RC:100B Reader unit it can store up to 24 hours of LAeq Time History data including Peak(C) and Overload exceedences. In addition to the new hardware, a suite of software is available to suport the new features and functions.

    Features

    • The unique noise dosimeter solution for personal noise measurements
    • Small size and light weight
    • New doseBadge with Time History data logging
    • New Reader unit with large backlit display and simple to use keys
    • No cables or controls reduces the risk of damage, misuse or tampering
    • Provides the parameters essential for Noise at Work measurements
    • Simple operation provides effective measurements
    • Operation & Download of doseBadge controlled via Infra-red link
    • Rechargable battery power for doseBadges
    • System can be expanded by adding doseBadges & Charging units
      Q=3 and Q=5 (OSHA) versions available
    • doseBadges can be programmed to either Q=3 (ISO) or Q=5 (OSHA)
    • Includes the new dBLink and doseBadge Database software.

    Applications

    • Measurement of Personal Noise Exposure
    • Assessing and Monitoring noise in the workplace
    • Ensuring compliance with Noise at Work regulations


      The doseBadge is a unique solution to the problems associated with the measurement of Personal Noise Exposure. The doseBadge itself has no controls or cables which means that the unit can be used in situations where the wearer is working in confined spaces or where there is a risk of a cable being caught in machinery.


     

    The doseBadge can only be operated (reset, calibrated, started and stopped) using the Reader unit or the Keyfob remote control. This dramatically reduces the possibility of the wearer tampering with the unit. The small size and light weight design of the doseBadge also reduces the risk of damage to the unit.

    The measurements provided the doseBadge have been designed to allow full compliance with Noise at Work regulations, while giving a simple output that can be easily understood.

    The design of the doseBadge allows the case to be much smaller than a conventional dosimeter. This ensures that the movement of the user is not constricted by cables and connectors. The unit has no controls or switches, with all functions controlled by either the RC:100B Reader unit, or by the RC:101B Keyfob Controller. The rechargable battery is sealed within the unit and is charged using the reader unit or by additional charging blocks.


    The RC:100B Reader can store up to 255 measurements, from any number of doseBadges. For example, the user can start a system with a single badge, a reader and a power supply. The system can then be expanded by simply adding more badges and block charging units as required. Kit cases are available to hold up to 10 badges and accessories.

    In most situations, the doseBadge is supplied as a complete measurement kit, with the user simply specifying the number of badges required. From August 2002, each reader unit is supplied with the dBLink and the doseBadge software and a download cable. The standard doseBadge measurement kits are listed below:

    doseBadge Measurement Kit
    Contains Charging Facility
    CK:100/1 1 x CR:100B doseBadge CU:100A 2 way Charger
    CK:100/2 2 x CR:100B doseBadges CU:100A 2 way Charger
    CK:100/5 5 x CR:100B doseBadges CU:101A 5 way Charger
    CK:100/10 10 x CR:100B doseBadges CU:102A 10 way Charger
    Specifications

    The CR:100B doseBadge Personal Noise Dosimeter and the RC:100B Reader Unit have been designed to meet the requirements of IEC 61252:1993 Personal Sound Exposure Meters and the ANSI S1.25:1991 Personal Noise Dosimeters. The CR:100B doseBadge and the RC:100B Reader unit must be used as a combination to ensure compliance with these standards.

    Applicable Standards

    IEC 61252:1993 Personal Sound Exposure Meters (See Note 1)
    ANSI S1.25:1991 Personal Noise Dosimeters Class Designation 2AS-90/80-5 (See Note 2)

    Note 1. When configured to the ISO Setting
    Note 2. When configured to the OSHA Setting

    Measurement Range

    80dB(A) to 130dB(A) Continuous
    To 140dB(C) Peak (factory configured to 140dB, 137dB or 134dB)

    Measurement Functions

    All versions doseBadge configuration, Calibration Record, Measurement Duration
    Peak Exceedence ,Overload Exceedence, Battery Status
    1 minute or 5 minute Time History (user selectable)

    The Time History is stored as LAeq along with a Peak(C) Exceedence and Overload in each period. The duration of the Time History is configured in the RC:100B Reader Unit.

    For ISO Q=3 LAeq, LEP,d, SEL, % Dose, Exposure (Pa2h), Estimated % Dose,
    Estimated Exposure (Pa2h)
    For OSHA Q=5 LAVG,TWA, % Dose, Estimated % Dose

    Memory

    The memory of the RC:100B Reader is 192kB. This allows the following number of measurements to be stored:

    No Time History Up to 500 Measurements
    With 24 hours of 1 minute Time History Up to 66 Measurements
    With 12 hours of 1 minute Time History Up to 128 Measurements
    With 8 hours of 1 minute Time History Up to 186 Measurements

    Weightings

    Frequency 'A' for all Sound Level measurements. 'C' for Peak Sound Pressure
    Time None as standard. 'S' (slow) for OSHA Version

    Exchange Rate

    Q=3 for ISO Version
    Q=5 for OSHA Version

    doseBadge Configuration

    Criterion Level User Selectable 80dB, 85dB, 87dB, 90dB
    Criterion Time 8 hours, 12 hours, 16 hours, 18 hours
    Threshold None (ISO), 80dB (OSHA)
    Peak Threshold Factory configured to 140dB. 137dB(C) or 134dB(C) Peak Option

    Power

    doseBadge Internal Nicad battery with intelligent charging system
    Reader 2 x AA/LR6

    Memory Backup

    The memory of the RC:100B Reader is protected when the main batteries are removed.

    Calibration

    Via internal Acoustic Calibrator to IEC 61942:1998 Class 1L
    Calibration Level 114dB
    Calibration Frequency 1kHz

    Output

    doseBadge Infrared to RC:100B Reader Unit
    Reader RS232 via Mini-Din

    Dimensions

    doseBadge Microphone Apex 12.7mm, Base 47mm, Height 30mmm
    RC:100B Reader 160mm x 80mm x 38mm

    Environmental

    Temperature -10C to +50C Operating
    -20C to +60C Storage
    Humidity Up to 99% RH Non Condensing

    Weight

    CR:100B 45 gms (1.6 oz)
    RC:100B Reader 400 gms (14 oz)

    Software

    The dBLink Download program and the doseBadge Database supplied as standard. Compatible with Microsoft Windows versions 95 or later. A free 9 pin RS232 Serial port is required on the PC for connection to the Reader. A USB Adaptor is available (ZL:101).

    Individual Components

    The individual components can be ordered separately, for example when expanding a kit with more doseBadges. The components are:

    CR:100B doseBadge
    RC:100A Reader Unit
    CU:195A Power Supply

    CU:100A 2 Way charger unit
    CU:101A 5 Way charger unit
    CU:102A 10 Way charger unit

    UA:100 Windshield for CR:100A doseBadge

    CK:100 Carrying Case
    SW:W100RD 100 Read for Windows Software
    ZL:700 RS232 Communications Cable

    RC:101A Keyfob Remote Control

    SP:024 Spare doseBadge Mounting Kit

    » «