| Sensor Probe |
U-51 |
U-52 |
U-52G |
U-53 |
U-53G |
| Max items displaying simultaneously |
9 |
10 |
11 |
10 |
11 |
| Measurement temperature |
-10 to -55°C |
| Max. sensor diameter |
Approx. 96 mm |
| Probe length |
Approx. 340 mm |
| Cable length |
2m, 10m, 30m |
| Mass |
Approx. 1800 g |
| Automatic calibration (use HORIBA 100-4) |
Applying to all models |
| Turbidity wiper |
– |
– |
– |
○ |
○ |
| Measurement depth |
Max. 30 m |
| Material contact with liquid |
PPS, glass, SUS316L, SUS304, FKM, PEEK, Titaniumu, FEP
membrane, POM |
| Water resistance |
JIS protection level 8 |
| Control Unit |
| Outer dimensions |
115(W) X 65(D) X 280(H) mm |
| Mass |
Approx. 800 g (Without Batteries) |
| LCD |
320 X 240 Graphic liquid crystal with backlight (black and
white) |
| Data memory |
10000 Data Sets |
| Communication |
USB peripheral |
| Battery |
C batteries x 4 ; CR2032 x 1 |
| Water resistance |
JIS protection level 7 (when sensor cable is
connected) |
| Battery life |
U-51, U-52, U-52G: Approx. 70 hours (without backlight) U-53, U-53G: Approx. 500 measurements |
| Storage temperature |
-10 to 60°C |
| Ambient temperature |
-5 to 45°C |
| pH |
| Measurement principle |
Glass electrode method |
| Range |
pH 0 to 14 |
| Resolution |
0.01pH |
| Repeatability |
±0.05pH |
| Accuracy |
±0.1pH |
| Conductivity |
| Measurement principle |
4 AC electrode method |
| Range |
0 to 10 S/m (0 to 100 mS/cm) |
| Resolution |
0.000 to 0.999 mS/cm: 0.001
1.00 to 9.99 mD/cm:
0.01
10.0 to 99.9 mS/cm: 0.1
0.0 to 99.9 mS/m: 0.1
0.100 to 0.999 S/m:
0.001
1.00 to 9.99 S/m: 0.01 |
| Repeatability |
±0.05% F.S. |
| Accuracy |
±1% F.S. (Median of two-point
calibration) |
| Seawater Specific Gravity |
| Measurement principle |
Conductivity conversion |
| Range |
0 to 50 δt |
| Resolution |
0.1 δt |
| Repeatability |
±2 δt |
| Accuracy |
±5 δt |
| Turbidity |
U-51 |
U-52 |
U-52G |
U-53 |
U-53G |
| Measurement principle |
– |
LED transmitting light source, 30° forward scattering
method |
Tungsten lamp transmitting source and 90° scattering
method |
| Range |
– |
0 to 800 NTU |
0 to 1000 NTU |
| Resolution |
– |
0.1 NTU |
0.01 NTU |
| Repeatability |
– |
±5% (reading) or ±0.5 NTU whichever is greater |
±3% (reading) or ±0.1 NTU whichever is greater |
| Accuracy |
– |
±5% (reading) or ±1 NTU whichever is greater |
0 to 10 NTU: ±0.5 NTU
10 to 1000 NTU: 3% (reading) or ±1
NTU whichever is greater |
| Dissolved Oxygen |
| Measurement principle |
Polarographic method |
| Range |
0 to 50.0 mg/L |
| Resolution |
0.01 mg/L |
| Repeatability |
±0.1 mg/L |
| Accuracy |
0 to 20 mg/L: ±0.2 mg/L, 20 to 50 mg/L: ±0.5
mg/L |
| Salinity |
| Measurement principle |
Conductivity conversion |
| Range |
0 to 70 PPT (permillage) |
| Resolution |
0.1 PPT |
| Repeatability |
±1 PPT |
| Accuracy |
±3 PPT |
| Total Dissolved Solids |
| Measurement principle |
Conductivity conversion |
| Range |
0 to 100 g/L |
| Resolution |
F.S. 0.1% |
| Repeatability |
±2 g/L |
| Accuracy |
±5 g/L |
| Temperature |
| Measurement principle |
Thermistor method |
| Range |
-10 to -55°C |
| Resolution |
0.01°C |
| Repeatability |
±0.10°C (at calibration point) |
| Accuracy |
JIS class B platinum thermometer sensor (±0.3 + 0.005 /
1°C) |
| Water Depth |
U-51 |
U-52 |
U-52G |
U-53 |
U-53G |
| Measurement principle |
– |
Pressure method |
| Range |
– |
0 to 30 m |
| Resolution |
– |
0.5 m |
| Repeatability |
– |
±1% F.S. |
| Accuracy |
– |
±0.3 m |
| Oxidation Reduction Potential |
| Measurement principle |
Platinum electrode method |
| Range |
-2000 mV to +2000 mV |
| Resolution |
1 mV |
| Repeatability |
±5 mV |
| Accuracy |
±15 mV |