The OCMA-310 is an automated portable unit for monitoring oil content in both fresh water and saltwater. The one-touch monitoring process is fully automated, from sample extraction to analysis and sample expulsion. The OCMA-310 utilizes an S-316 extraction solvent to ensure superior oil particle stability and infrared absorptiometry for sample analysis. Using 0mg/l -200mg/l dynamic ranging, the unit is capable of analyzing both high and low concentrations. The OCMA-310 provides swift, highly accurate analysis of the oil content of both fresh water and saltwater as well as that of such fluids as those found at factories and wastewater treatment plants.

Horiba OCMA-310


A microcomputer provides automated analysis of the samples at the touch of a button.

Dynamic ranging provides analysis from 0mg/l-200mg/l without manual switching.

The use of an S-316 extraction solvent ensures the highest levels of oil particle stability.

A digital display eliminates the misreading of analysis results. A visual display alerts users to the unit's operating status.

The use of a proprietary syringe ensures a fixed amount of sample water and solvent.

Environment

Surveying water quality and hazardous waste sites

Industry

Monitoring wastewater influent and effluent

Marine transportation

Checking bilge and ballast discharge

Petroleum processing

Monitoring the efficiency of oil/water separation processes

Oil depots

Monitoring the discharge produced when cleaning storage tanks

Automotive

Monitoring waste-water discharge from service stations

Quality Assurance

Measuring residual oil on textiles and metal parts

Measurement principle:

Solvent extraction, NDIR
analysis (Infrared spectroscopy)

Measurement range:

0mg/l ~ 200mg/l dynamic ranging

Resolution:

0 to 99.9 mg/L; 0.1 mg/L
100 to 200 mg/L; 1 mg/L

Repeatability:

0 to 9.9 mg/L; ± 0.2 mg/L ± 1 digit
10.0 to 99.9 mg/L; ± 2 mg/L ± 1 digit
100 to 200 mg/L; ± 4 mg/L ± 1 digit

Measurement:

Manually controlled

Calibration:

One touch calibration after the calibration standard is introduced to the instrument

Extraction Solvent:

S-316 solvent

Sample/Solvent Volume:

20 mL, ratio of sample to solvent is 2:1.

Extraction Method:

Built-in extractor for oily water samples

Display:

Measured value; 3 digits LCD with backlight
Message; Character display LCD with back-light (16×2 characters)

Functions:

Data memory up to 50 data, Self diagnostics, Stable type display, Interactive operation, Calendar clock

Output:

RS-232C, parallel printer port

Option: Analog output 0 to 200 mg/L;0 to 1 V

Ambient Temperature:

0 to 40 ℃

Power:

100 to 120 V AC ±10%,50/60 Hz, 80 VA
200 to 240 V AC ±10%,50/60 Hz, 120 VA

Dimensions:

200(W)×315(D)×342(H) mm

Mass:

Approx. 7 kg

Title
Description
Download
Horiba OCMA-310 Datasheet
Datasheet
PDF Icon
Enviro-Equipment
Specials || Used Equipment
Rental Options &
Delivery Details
Service/Repair & Maint.
Ask The Expert! || Blog
Sign Up For Our Newsletter!
Why Buy From Us?
Frequently Asked Questions
Locations & Contact Information
Account Setup

check out our Facebook
follow us on Twitter
see our videos on YouTube

Share This Website!

Click to Return to MAIN Home Page
Click to Return to MAIN Home Page
Click to Return to MAIN Home Page
Click to Return to MAIN Home Page
home || about us || products - general: rentalsales || products - remediation: rentalsalesused || services || order form || contact information || site map
© 2002-2010 — All Rights Reserved — Enviro-Equipment, Inc — Environmental Equipment & Supplies: Rental, Sales, Service & Repair