Technical Information Intrinsically Safe Models |
Flow range
Flow control accuracy
Flow performance
(Flow, max vacuum pressure) |
|
0.8 to 4.0 liters/minute standard flow rate
(5 to 750 milliliters/minute with Low Flow adaptor) <±3.0% for selected flow digitally controlled for utmost stability
1.0 liter/minute @ 32 in H2O (80 cm H2O)
2.0 liters/minute @ 22 in H2O (55 cm H2O)
3.5 liters/minute @ 6.4 in H2O (16 cm H2O) |
| Battery voltage and capacity |
|
4.8 V NiMH / 1.7 Ah |
| Inlet pulsation ratio |
|
<10% using cyclone
@ 2.0 liters/minute approx. 3.0 in H2O
Satisfies NIOSH 0600 and EN 1232 |
| Displayed values |
|
Real time flow rate, volume sampled, elapsed sample time,
temperature, operating mode, program details |
| Pump MTBF (service interval) |
|
Typically >3000 hours |
Operating temperature
Storage temperature |
|
41 to 104 °F (5 to 40 °C)
50 to 122 ° F (10 to 50 °C) |
Charging technique
Typical charge time
External Power Supply Unit
(for drop in charger) |
|
Drop in intelligent fast charger
Fast charge with safety time out takes 4.5 hours to fully charge pump
Casella 182052B single pump fast charger or 182226A 5 way pump
fast charger keeps pump(s) always ready for use |
Memory storage of data
Memory protection time |
|
100 events saved in Apex Pro IS model
Approximately 20 minutes (with main battery removed) |
| Communications |
|
Infra red (via RS232 IR transducer) to Pump Manager software |
| Recorded values |
|
Start and stop times, flow rate, average sample temperature,
total volume sampled (in l or m3) |
Physical characteristics - Size
Weight (including. battery) |
|
5.8 x 3.4 x 1.8 in (145 x 85 x 46 mm)
17.8 oz (500 gm) |
| Intrinsic Safety Approvals |
|
ANSI/UL Class I, Divisions 1 & 2, Groups A, B, C & D; Class II,
Divisions 1 & 2, Groups E, F & G; Class III hazardous locations
ATEX: 100-94/9/EC Directive EEx ia IIC T4, Equipment group II
Category 2 G Gas |
 |
| Pump Performance |
Typical filter pressure load in inches of H2O, typical battery life in hours |
Filter type |
1.0 liter/min |
2.0 liters/min |
2.2 liters/min |
3.5 liters/min |
25 mm GFA |
2.0 in, 34 hrs |
3.5 in, 18 hrs |
3.7 in, 17 hrs |
6.3 in, 10 hrs |
| 25 mm 0.8 µm |
7.0 in, 18 hrs |
12.2 in, 10 hrs |
13.4 in, 9 hrs |
N/A |
| 25 mm 1.2 µm |
4.3 in, 24 hrs |
8.3 in, 13 hrs |
9.0 in, 11 hrs |
N/A |
| 37 mm GFA |
1.2 in, 41 hrs |
2.0 in, 22 hrs |
2.4 in, 19 hrs |
3.5 in, 12 hrs |
| 37 mm 0.8 µm |
2.6 in, 30 hrs |
5.1 in, 16 hrs |
5.5 in, 15 hrs |
N/A |
 |
| Ordering information |
| 182050B |
|
Apex Standard IS pump with NiMH battery pack (less charger unit) |
| 182060B |
|
Apex Pro IS pump with NiMH Battery pack (less charger unit) |
| 182050B/K1 |
|
Apex Standard IS pump including charger unit |
| 182060B/K1 |
|
Apex Pro IS pump including charger unit |
| 182050B/K5 |
|
Apex Standard IS kit (5 pumps) including 5 way charger and case |
| 182060B/K5 |
|
Apex Pro IS kit (5 pumps) including 5 way charger, sware and case |
| 182091A |
|
MS Windows based Pump Manager software with IR link and cable |
| 182261B |
|
Single battery charger for Apex IS & non-IS pumps (not UL approved) |
| 182226A |
|
5 way battery charger for Apex IS & non-IS pumps (not UL approved) |
| 182013C |
|